| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 6000.00 | 2.189 | 2.088 | 
| 3000.00 | 1.582 | 1.481 | 
| 1500.00 | 1.243 | 1.142 | 
| 750.00 | 0.935 | 0.834 | 
| 375.00 | 0.503 | 0.402 | 
| 187.50 | 0.285 | 0.184 | 
| 93.75 | 0.219 | 0.118 | 
| 0.00 | 0.101 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-97% | 89% | 
| EDTA plasma(n=5) | 85-99% | 92% | 
| Heparin plasma(n=5) | 90-105% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 79-96% | 85-99% | 79-93% | 89-97% | 
| EDTA plasma(n=5) | 88-105% | 86-92% | 81-93% | 79-97% | 
| Heparin plasma(n=5) | 85-102% | 87-96% | 92-101% | 89-104% | 