| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µIU/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.964 | 1.875 |
| 500.00 | 1.477 | 1.388 |
| 250.00 | 1.116 | 1.027 |
| 125.00 | 0.668 | 0.579 |
| 62.50 | 0.490 | 0.401 |
| 31.25 | 0.309 | 0.220 |
| 15.63 | 0.196 | 0.107 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-102% | 95% |
| EDTA plasma(n=5) | 85-109% | 97% |
| Heparin plasma(n=5) | 90-103% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-108% | 86-106% | 88-113% | 89-112% |
| EDTA plasma(n=5) | 82-102% | 90-113% | 84-111% | 86-109% |
| Heparin plasma(n=5) | 86-112% | 84-97% | 90-97% | 92-98% |