| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 0.182 | |
| 50.00 | 0.336 | |
| 25.00 | 0.557 | |
| 12.50 | 0.993 | |
| 6.25 | 1.139 | |
| 3.13 | 1.565 | |
| 1.57 | 1.762 | |
| 0.00 | 2.112 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-92% | 85% | 
| EDTA plasma(n=5) | 88-106% | 97% | 
| Heparin plasma(n=5) | 86-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 84-99% | 81-95% | 98-107% | 96-101% | 
| EDTA plasma(n=5) | 88-95% | 98-107% | 90-99% | 88-97% | 
| Heparin plasma(n=5) | 93-101% | 97-103% | 91-99% | 90-98% |