| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.345 | 2.249 | 
| 2500.00 | 1.761 | 1.665 | 
| 1250.00 | 1.183 | 1.087 | 
| 625.00 | 0.647 | 0.551 | 
| 312.50 | 0.521 | 0.425 | 
| 156.25 | 0.393 | 0.297 | 
| 78.13 | 0.258 | 0.162 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-90% | 84% | 
| EDTA plasma(n=5) | 84-97% | 91% | 
| Heparin plasma(n=5) | 89-105% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-102% | 92-101% | 88-102% | 87-102% | 
| EDTA plasma(n=5) | 88-97% | 96-105% | 81-96% | 93-107% | 
| Heparin plasma(n=5) | 85-94% | 87-96% | 92-101% | 82-95% | 