| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.035 | 1.940 |
| 2500.00 | 1.617 | 1.522 |
| 1250.00 | 1.196 | 1.101 |
| 625.00 | 0.895 | 0.800 |
| 312.50 | 0.558 | 0.463 |
| 156.25 | 0.379 | 0.284 |
| 78.13 | 0.212 | 0.117 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 90% |
| EDTA plasma(n=5) | 87-95% | 91% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-103% | 97-105% | 79-90% | 87-96% |
| EDTA plasma(n=5) | 93-99% | 87-101% | 85-95% | 85-96% |
| Heparin plasma(n=5) | 88-100% | 95-103% | 89-97% | 85-94% |