| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.052 | 1.949 | 
| 500.00 | 1.509 | 1.406 | 
| 250.00 | 1.285 | 1.182 | 
| 125.00 | 0.765 | 0.662 | 
| 62.50 | 0.493 | 0.390 | 
| 31.25 | 0.312 | 0.209 | 
| 15.63 | 0.186 | 0.083 | 
| 0.00 | 0.103 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-102% | 95% | 
| EDTA plasma(n=5) | 79-97% | 88% | 
| Heparin plasma(n=5) | 83-95% | 89% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-94% | 79-96% | 85-99% | 
| EDTA plasma(n=5) | 88-101% | 86-92% | 83-98% | 96-103% | 
| Heparin plasma(n=5) | 85-94% | 87-103% | 81-95% | 82-90% | 