| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 50.00 | 2.231 | 2.149 | 
| 25.00 | 1.487 | 1.405 | 
| 12.50 | 1.069 | 0.987 | 
| 6.25 | 0.874 | 0.792 | 
| 3.13 | 0.479 | 0.397 | 
| 1.57 | 0.384 | 0.302 | 
| 0.79 | 0.248 | 0.166 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-103% | 99% | 
| EDTA plasma(n=5) | 92-102% | 97% | 
| Heparin plasma(n=5) | 81-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-96% | 93-104% | 89-97% | 85-94% | 
| EDTA plasma(n=5) | 83-96% | 86-95% | 90-102% | 81-93% | 
| Heparin plasma(n=5) | 83-94% | 85-96% | 92-103% | 86-97% | 