| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.996 | 1.914 |
| 1000.00 | 1.588 | 1.506 |
| 500.00 | 1.185 | 1.103 |
| 250.00 | 0.838 | 0.756 |
| 125.00 | 0.608 | 0.526 |
| 62.50 | 0.324 | 0.242 |
| 31.25 | 0.211 | 0.129 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-99% | 92% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 93-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-92% | 86-97% | 90-101% | 92-102% |
| EDTA plasma(n=5) | 89-97% | 95-105% | 87-102% | 92-102% |
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 79-86% |