| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.121 | 2.034 |
| 2000.00 | 1.687 | 1.600 |
| 1000.00 | 1.182 | 1.095 |
| 500.00 | 0.871 | 0.784 |
| 250.00 | 0.539 | 0.452 |
| 125.00 | 0.313 | 0.226 |
| 62.50 | 0.241 | 0.154 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 92% |
| EDTA plasma(n=5) | 79-97% | 88% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-96% | 85-96% | 87-101% | 96-105% |
| EDTA plasma(n=5) | 90-96% | 95-102% | 88-95% | 92-101% |
| Heparin plasma(n=5) | 88-95% | 92-101% | 87-98% | 85-96% |