| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 0.191 | |
| 150.00 | 0.356 | |
| 75.00 | 0.524 | |
| 37.50 | 0.897 | |
| 18.75 | 1.152 | |
| 9.38 | 1.522 | |
| 4.69 | 1.789 | |
| 0.00 | 2.213 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-103% | 87% |
| EDTA plasma(n=5) | 93-105% | 99% |
| Heparin plasma(n=5) | 84-93% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |
| EDTA plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |
| Heparin plasma(n=5) | 83-97% | 82-96% | 86-98% | 93-101% |