| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 8000.00 | 0.235 | |
| 4000.00 | 0.389 | |
| 2000.00 | 0.672 | |
| 1000.00 | 0.913 | |
| 500.00 | 1.197 | |
| 250.00 | 1.565 | |
| 125.00 | 1.793 | |
| 0.00 | 2.325 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-103% | 94% |
| EDTA plasma(n=5) | 95-105% | 100% |
| Heparin plasma(n=5) | 93-101% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-101% | 85-97% |
| EDTA plasma(n=5) | 86-100% | 93-108% | 80-101% | 87-96% |
| Heparin plasma(n=5) | 86-93% | 88-92% | 86-97% | 90-101% |