| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 0.185 | |
| 500.00 | 0.397 | |
| 250.00 | 0.611 | |
| 125.00 | 0.993 | |
| 62.50 | 1.139 | |
| 31.25 | 1.583 | |
| 15.63 | 1.761 | |
| 0.00 | 2.211 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 89-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 95-103% | 98-106% |
| EDTA plasma(n=5) | 85-94% | 87-96% | 82-95% | 81-93% |
| Heparin plasma(n=5) | 87-94% | 85-96% | 92-102% | 81-90% |