Products

Human HS(Heparan Sulfate) ELISA Kit
One-click to copy product information
ELK0768
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Human HS(Heparan Sulfate) ELISA Kit
Reactivity: Human
Alternative Names: HS; Heparan Sulfate
Assay Type: Competitive Inhibition
Sensitivity: 51.9 ng/mL
Standard: 8000 ng/mL
Detection Range: 125-8000 ng/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 2h
Research Area: Signal transduction;Metabolic pathway;Tumor immunity;Infection immunity;Developmental science;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human HS. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human HS. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human HS in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
8000.00 0.232
4000.00 0.413
2000.00 0.596
1000.00 0.997
500.00 1.236
250.00 1.592
125.00 1.813
0.00 2.226

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant HS and the recovery rates were calculated by comparing the measured value to the expected amount of HS in samples.
Matrix Recovery range Average
serum(n=5) 84-106% 100%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 87-93% 90%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HS and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 79-91% 96-105% 80-92% 91-103%
EDTA plasma(n=5) 92-104% 87-96% 89-101% 82-90%
Heparin plasma(n=5) 84-99% 83-101% 86-99% 93-101%