| 48 T | $370.00 |
| 96 T | $528.00 |
| 96 T*5 | $2244.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 20000.00 | 0.217 | |
| 10000.00 | 0.467 | |
| 5000.00 | 0.722 | |
| 2500.00 | 0.993 | |
| 1250.00 | 1.258 | |
| 625.00 | 1.589 | |
| 312.50 | 1.733 | |
| 0.00 | 2.256 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 88-103% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 93-107% | 93-102% | 91-105% |
| EDTA plasma(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |
| Heparin plasma(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |