| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.952 | 1.875 |
| 2500.00 | 1.647 | 1.570 |
| 1250.00 | 1.183 | 1.106 |
| 625.00 | 0.792 | 0.715 |
| 312.50 | 0.539 | 0.462 |
| 156.25 | 0.381 | 0.304 |
| 78.13 | 0.152 | 0.075 |
| 0.00 | 0.077 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 82-98% | 90% |
| Heparin plasma(n=5) | 90-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-106% | 85-104% | 90-98% | 82-98% |
| EDTA plasma(n=5) | 86-97% | 91-102% | 81-97% | 93-107% |
| Heparin plasma(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |