| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.074 | 1.989 |
| 500.00 | 1.642 | 1.557 |
| 250.00 | 1.069 | 0.984 |
| 125.00 | 0.801 | 0.716 |
| 62.50 | 0.474 | 0.389 |
| 31.25 | 0.342 | 0.257 |
| 15.63 | 0.138 | 0.053 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 87-93% | 90% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 89-103% | 97-105% | 91-99% |
| EDTA plasma(n=5) | 88-96% | 86-95% | 90-103% | 87-90% |
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |