| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.120 | 1.977 |
| 50.00 | 1.519 | 1.376 |
| 25.00 | 1.194 | 1.051 |
| 12.50 | 0.985 | 0.842 |
| 6.25 | 0.645 | 0.502 |
| 3.13 | 0.370 | 0.227 |
| 1.56 | 0.236 | 0.093 |
| 0.00 | 0.143 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-117% | 99% |
| EDTA plasma(n=5) | 89-97% | 93% |
| Heparin plasma(n=5) | 80-108% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-112% | 87-98% | 83-94% | 83-120% |
| EDTA plasma(n=5) | 93-99% | 83-115% | 88-94% | 82-118% |
| Heparin plasma(n=5) | 86-120% | 81-102% | 86-115% | 85-116% |