| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.035 | 1.937 |
| 50.00 | 1.636 | 1.538 |
| 25.00 | 1.243 | 1.145 |
| 12.50 | 0.896 | 0.798 |
| 6.25 | 0.543 | 0.445 |
| 3.13 | 0.386 | 0.288 |
| 1.57 | 0.226 | 0.128 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-101% | 98% |
| EDTA plasma(n=5) | 93-104% | 99% |
| Heparin plasma(n=5) | 85-94% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 80-87% | 79-87% | 79-92% | 92-101% |
| EDTA plasma(n=5) | 87-100% | 78-91% | 78-87% | 77-90% |
| Heparin plasma(n=5) | 96-103% | 86-93% | 86-96% | 86-92% |