| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.951 | 1.869 |
| 5.00 | 1.700 | 1.618 |
| 2.50 | 1.128 | 1.046 |
| 1.25 | 0.834 | 0.752 |
| 0.63 | 0.516 | 0.434 |
| 0.32 | 0.364 | 0.282 |
| 0.16 | 0.134 | 0.052 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 79-96% | 92-101% | 83-90% |
| EDTA plasma(n=5) | 95-104% | 85-96% | 87-102% | 91-99% |
| Heparin plasma(n=5) | 89-99% | 83-95% | 98-104% | 96-106% |