| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.075 | 1.977 |
| 1000.00 | 1.685 | 1.587 |
| 500.00 | 1.142 | 1.044 |
| 250.00 | 0.806 | 0.708 |
| 125.00 | 0.508 | 0.410 |
| 62.50 | 0.357 | 0.259 |
| 31.25 | 0.256 | 0.158 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-94% | 97-105% | 88-101% | 82-96% |
| EDTA plasma(n=5) | 83-97% | 92-103% | 87-98% | 93-101% |
| Heparin plasma(n=5) | 83-101% | 90-99% | 87-103% | 89-97% |