| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 30.00 | 2.045 | 1.953 |
| 15.00 | 1.612 | 1.520 |
| 7.50 | 1.182 | 1.090 |
| 3.75 | 0.861 | 0.769 |
| 1.88 | 0.567 | 0.475 |
| 0.94 | 0.324 | 0.232 |
| 0.47 | 0.215 | 0.123 |
| 0.00 | 0.092 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-105% | 98% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 79-92% | 82-95% | 86-97% |
| EDTA plasma(n=5) | 79-92% | 85-96% | 80-98% | 95-102% |
| Heparin plasma(n=5) | 95-103% | 86-92% | 91-102% | 89-97% |