| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.201 | 2.120 |
| 2000.00 | 1.715 | 1.634 |
| 1000.00 | 1.027 | 0.946 |
| 500.00 | 0.869 | 0.788 |
| 250.00 | 0.543 | 0.462 |
| 125.00 | 0.362 | 0.281 |
| 62.50 | 0.156 | 0.075 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-96% | 87% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-93% | 86-98% | 91-103% | 82-95% |
| EDTA plasma(n=5) | 89-99% | 83-96% | 98-104% | 88-101% |
| Heparin plasma(n=5) | 85-102% | 84-99% | 97-101% | 89-102% |