| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.865 | 1.779 |
| 1000.00 | 1.543 | 1.457 |
| 500.00 | 1.134 | 1.048 |
| 250.00 | 0.850 | 0.764 |
| 125.00 | 0.568 | 0.482 |
| 62.50 | 0.237 | 0.151 |
| 31.25 | 0.185 | 0.099 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-119% | 101% |
| EDTA plasma(n=5) | 94-119% | 107% |
| Heparin plasma(n=5) | 94-99% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-115% | 92-119% | 85-112% | 84-115% |
| EDTA plasma(n=5) | 81-95% | 84-106% | 86-115% | 89-100% |
| Heparin plasma(n=5) | 90-97% | 86-100% | 81-100% | 82-115% |