| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 6000.00 | 2.132 | 2.045 |
| 3000.00 | 1.624 | 1.537 |
| 1500.00 | 1.223 | 1.136 |
| 750.00 | 0.933 | 0.846 |
| 375.00 | 0.593 | 0.506 |
| 187.50 | 0.357 | 0.270 |
| 93.75 | 0.189 | 0.102 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 83-97% | 90% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-104% | 93-102% | 82-97% | 92-101% |
| EDTA plasma(n=5) | 95-104% | 85-96% | 87-102% | 91-99% |
| Heparin plasma(n=5) | 91-97% | 86-92% | 79-93% | 89-106% |