| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.105 | 2.008 |
| 2500.00 | 1.403 | 1.306 |
| 1250.00 | 1.120 | 1.023 |
| 625.00 | 0.876 | 0.779 |
| 312.50 | 0.560 | 0.463 |
| 156.25 | 0.296 | 0.199 |
| 78.13 | 0.230 | 0.133 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-100% | 96% |
| EDTA plasma(n=5) | 88-107% | 98% |
| Heparin plasma(n=5) | 90-100% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-112% | 85-91% | 91-106% | 93-105% |
| EDTA plasma(n=5) | 92-106% | 94-116% | 87-119% | 86-111% |
| Heparin plasma(n=5) | 80-119% | 83-117% | 81-91% | 90-110% |