| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.978 | 1.883 |
| 1000.00 | 1.636 | 1.541 |
| 500.00 | 1.135 | 1.040 |
| 250.00 | 0.775 | 0.680 |
| 125.00 | 0.487 | 0.392 |
| 62.50 | 0.322 | 0.227 |
| 31.25 | 0.238 | 0.143 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 78-90% | 83-95% | 86-97% |
| EDTA plasma(n=5) | 95-102% | 96-104% | 87-98% | 95-107% |
| Heparin plasma(n=5) | 78-97% | 86-97% | 81-103% | 88-97% |