| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 0.179 | |
| 250.00 | 0.396 | |
| 125.00 | 0.692 | |
| 62.50 | 0.823 | |
| 31.25 | 1.258 | |
| 15.63 | 1.572 | |
| 7.82 | 1.697 | |
| 0.00 | 2.219 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-98% | 90% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 88-102% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-90% | 87-98% | 85-96% | 85-95% |
| EDTA plasma(n=5) | 87-96% | 92-105% | 89-102% | 92-101% |
| Heparin plasma(n=5) | 91-98% | 79-93% | 92-101% | 82-90% |