| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 0.179 | |
| 150.00 | 0.396 | |
| 75.00 | 0.551 | |
| 37.50 | 0.857 | |
| 18.75 | 1.112 | |
| 9.38 | 1.457 | |
| 4.69 | 1.793 | |
| 0.00 | 2.226 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-98% | 90% |
| EDTA plasma(n=5) | 84-98% | 91% |
| Heparin plasma(n=5) | 80-96% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-92% | 92-96% | 86-98% | 95-101% |
| EDTA plasma(n=5) | 87-93% | 85-95% | 89-99% | 85-95% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |