| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.255 | 2.173 |
| 2500.00 | 1.703 | 1.621 |
| 1250.00 | 1.116 | 1.034 |
| 625.00 | 0.896 | 0.814 |
| 312.50 | 0.532 | 0.450 |
| 156.25 | 0.345 | 0.263 |
| 78.13 | 0.225 | 0.143 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 90-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-93% | 87-96% | 79-91% | 95-103% |
| EDTA plasma(n=5) | 86-98% | 93-106% | 89-97% | 85-94% |
| Heparin plasma(n=5) | 93-102% | 87-96% | 85-98% | 91-103% |