| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.946 | 1.875 |
| 5.00 | 1.573 | 1.502 |
| 2.50 | 1.147 | 1.076 |
| 1.25 | 0.865 | 0.794 |
| 0.63 | 0.623 | 0.552 |
| 0.32 | 0.326 | 0.255 |
| 0.16 | 0.251 | 0.180 |
| 0.00 | 0.071 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-95% | 86% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 78-90% | 83-95% | 86-97% |
| EDTA plasma(n=5) | 95-102% | 96-104% | 87-98% | 95-107% |
| Heparin plasma(n=5) | 78-97% | 86-97% | 81-103% | 88-97% |