| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.201 | 2.113 |
| 500.00 | 1.565 | 1.477 |
| 250.00 | 1.115 | 1.027 |
| 125.00 | 0.928 | 0.840 |
| 62.50 | 0.582 | 0.494 |
| 31.25 | 0.372 | 0.284 |
| 15.63 | 0.196 | 0.108 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 78-93% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 87-96% | 85-96% | 85-95% |
| EDTA plasma(n=5) | 79-91% | 89-103% | 82-95% | 86-99% |
| Heparin plasma(n=5) | 91-102% | 86-102% | 87-99% | 99-105% |