| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.998 | 1.912 |
| 5.00 | 1.624 | 1.538 |
| 2.50 | 0.976 | 0.890 |
| 1.25 | 0.728 | 0.642 |
| 0.63 | 0.508 | 0.422 |
| 0.32 | 0.357 | 0.271 |
| 0.16 | 0.319 | 0.233 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 78-90% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-103% | 98-105% | 96-101% |
| EDTA plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |
| Heparin plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |