| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.052 | 1.965 |
| 5.00 | 1.717 | 1.630 |
| 2.50 | 1.282 | 1.195 |
| 1.25 | 0.797 | 0.710 |
| 0.63 | 0.535 | 0.448 |
| 0.32 | 0.324 | 0.237 |
| 0.16 | 0.175 | 0.088 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 90% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-101% | 97-103% | 91-99% | 90-98% |
| EDTA plasma(n=5) | 92-106% | 92-101% | 80-95% | 87-96% |
| Heparin plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |