| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 3000.00 | 2.059 | 1.979 |
| 1500.00 | 1.647 | 1.567 |
| 750.00 | 1.185 | 1.105 |
| 375.00 | 0.752 | 0.672 |
| 187.50 | 0.539 | 0.459 |
| 93.75 | 0.363 | 0.283 |
| 46.88 | 0.206 | 0.126 |
| 0.00 | 0.080 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-104% | 98% |
| EDTA plasma(n=5) | 82-90% | 86% |
| Heparin plasma(n=5) | 84-92% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-93% | 96-105% | 87-98% | 85-97% |
| EDTA plasma(n=5) | 92-106% | 92-101% | 80-95% | 87-96% |
| Heparin plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |