| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.079 | 1.994 |
| 5.00 | 1.643 | 1.558 |
| 2.50 | 1.197 | 1.112 |
| 1.25 | 0.828 | 0.743 |
| 0.63 | 0.516 | 0.431 |
| 0.32 | 0.357 | 0.272 |
| 0.16 | 0.186 | 0.101 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 97-105% | 101% |
| EDTA plasma(n=5) | 83-99% | 95% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-96% | 92-105% | 89-102% | 92-101% |
| EDTA plasma(n=5) | 97-105% | 84-96% | 87-98% | 95-102% |
| Heparin plasma(n=5) | 79-97% | 90-99% | 95-103% | 89-97% |