| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.209 | 2.107 |
| 2500.00 | 1.595 | 1.493 |
| 1250.00 | 1.172 | 1.070 |
| 625.00 | 0.885 | 0.783 |
| 312.50 | 0.579 | 0.477 |
| 156.25 | 0.332 | 0.230 |
| 78.13 | 0.182 | 0.080 |
| 0.00 | 0.102 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-115% | 105% |
| EDTA plasma(n=5) | 89-95% | 92% |
| Heparin plasma(n=5) | 91-110% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-117% | 84-114% | 87-119% | 81-109% |
| EDTA plasma(n=5) | 92-109% | 83-108% | 92-104% | 81-88% |
| Heparin plasma(n=5) | 88-112% | 84-108% | 93-117% | 90-100% |