| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.368 | 2.282 |
| 500.00 | 1.563 | 1.477 |
| 250.00 | 1.123 | 1.037 |
| 125.00 | 0.726 | 0.640 |
| 62.50 | 0.596 | 0.510 |
| 31.25 | 0.387 | 0.301 |
| 15.63 | 0.232 | 0.146 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 90-98% | 80-93% | 92-105% |
| EDTA plasma(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |
| Heparin plasma(n=5) | 87-94% | 85-96% | 89-100% | 85-96% |