| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.979 | 1.893 |
| 1000.00 | 1.608 | 1.522 |
| 500.00 | 1.142 | 1.056 |
| 250.00 | 0.891 | 0.805 |
| 125.00 | 0.462 | 0.376 |
| 62.50 | 0.279 | 0.193 |
| 31.25 | 0.206 | 0.120 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 80-93% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-105% | 83-92% | 89-98% | 85-94% |
| EDTA plasma(n=5) | 89-98% | 83-103% | 98-105% | 92-101% |
| Heparin plasma(n=5) | 85-105% | 87-98% | 91-102% | 86-102% |