| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.179 | 2.097 |
| 2500.00 | 1.579 | 1.497 |
| 1250.00 | 1.139 | 1.057 |
| 625.00 | 0.797 | 0.715 |
| 312.50 | 0.529 | 0.447 |
| 156.25 | 0.327 | 0.245 |
| 78.13 | 0.187 | 0.105 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 88-96% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-102% | 83-96% | 95-102% | 87-98% |
| EDTA plasma(n=5) | 86-100% | 93-108% | 80-101% | 87-96% |
| Heparin plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |