| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.332 | 2.246 |
| 2500.00 | 1.720 | 1.634 |
| 1250.00 | 1.088 | 1.002 |
| 625.00 | 0.826 | 0.740 |
| 312.50 | 0.509 | 0.423 |
| 156.25 | 0.373 | 0.287 |
| 78.13 | 0.189 | 0.103 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 89-103% | 96% |
| Heparin plasma(n=5) | 78-93% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-93% | 86-98% | 91-103% | 82-95% |
| EDTA plasma(n=5) | 87-96% | 91-101% | 87-98% | 89-103% |
| Heparin plasma(n=5) | 84-94% | 85-103% | 94-105% | 92-101% |