 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.154 | 2.072 | 
| 1000.00 | 1.534 | 1.452 | 
| 500.00 | 1.135 | 1.053 | 
| 250.00 | 0.847 | 0.765 | 
| 125.00 | 0.535 | 0.453 | 
| 62.50 | 0.386 | 0.304 | 
| 31.25 | 0.175 | 0.093 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-96% | 88% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 83-99% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-96% | 85-97% | 85-94% | 87-96% | 
| EDTA plasma(n=5) | 93-102% | 93-106% | 87-101% | 91-104% | 
| Heparin plasma(n=5) | 89-97% | 93-102% | 81-96% | 93-106% | 
