| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 0.189 | |
| 500.00 | 0.386 | |
| 250.00 | 0.611 | |
| 125.00 | 0.823 | |
| 62.50 | 1.169 | |
| 31.25 | 1.592 | |
| 15.63 | 1.697 | |
| 0.00 | 2.213 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-94% | 87% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 90-98% | 80-93% | 92-105% |
| EDTA plasma(n=5) | 87-98% | 85-92% | 79-96% | 89-100% |
| Heparin plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |