| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.932 | 1.849 |
| 1000.00 | 1.674 | 1.591 |
| 500.00 | 1.174 | 1.091 |
| 250.00 | 0.828 | 0.745 |
| 125.00 | 0.556 | 0.473 |
| 62.50 | 0.351 | 0.268 |
| 31.25 | 0.176 | 0.093 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-93% | 87-96% | 79-91% | 95-103% |
| EDTA plasma(n=5) | 85-94% | 92-101% | 87-96% | 82-90% |
| Heparin plasma(n=5) | 93-105% | 88-102% | 87-98% | 85-96% |