| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.179 | 2.095 |
| 500.00 | 1.636 | 1.552 |
| 250.00 | 1.175 | 1.091 |
| 125.00 | 0.895 | 0.811 |
| 62.50 | 0.547 | 0.463 |
| 31.25 | 0.388 | 0.304 |
| 15.63 | 0.256 | 0.172 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 88-102% | 95% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-99% | 83-89% | 88-94% | 87-93% |
| EDTA plasma(n=5) | 89-101% | 85-96% | 95-104% | 82-90% |
| Heparin plasma(n=5) | 95-104% | 97-106% | 85-98% | 81-92% |