| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.952 | 1.849 |
| 2500.00 | 1.529 | 1.426 |
| 1250.00 | 1.027 | 0.924 |
| 625.00 | 0.866 | 0.763 |
| 312.50 | 0.559 | 0.456 |
| 156.25 | 0.377 | 0.274 |
| 78.13 | 0.265 | 0.162 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-109% | 102% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-101% | 89-99% | 85-95% |
| EDTA plasma(n=5) | 90-99% | 87-103% | 78-91% | 83-97% |
| Heparin plasma(n=5) | 85-102% | 83-96% | 95-102% | 87-98% |