| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.054 | 1.968 |
| 5.00 | 1.630 | 1.544 |
| 2.50 | 1.251 | 1.165 |
| 1.25 | 0.831 | 0.745 |
| 0.63 | 0.496 | 0.410 |
| 0.32 | 0.331 | 0.245 |
| 0.16 | 0.325 | 0.239 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-97% | 90% |
| EDTA plasma(n=5) | 82-94% | 88% |
| Heparin plasma(n=5) | 81-97% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-100% | 84-98% | 95-101% | 88-102% |
| EDTA plasma(n=5) | 90-103% | 97-106% | 82-90% | 92-101% |
| Heparin plasma(n=5) | 85-92% | 82-96% | 87-105% | 87-92% |