| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.266 | 2.162 |
| 500.00 | 1.699 | 1.595 |
| 250.00 | 1.132 | 1.028 |
| 125.00 | 0.895 | 0.791 |
| 62.50 | 0.525 | 0.421 |
| 31.25 | 0.354 | 0.250 |
| 15.63 | 0.230 | 0.126 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 82-98% | 90% |
| Heparin plasma(n=5) | 90-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |
| EDTA plasma(n=5) | 87-99% | 85-96% | 87-101% | 89-102% |
| Heparin plasma(n=5) | 93-102% | 90-99% | 95-106% | 87-101% |