| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.164 | 2.066 |
| 500.00 | 1.687 | 1.589 |
| 250.00 | 1.117 | 1.019 |
| 125.00 | 0.819 | 0.721 |
| 62.50 | 0.523 | 0.425 |
| 31.25 | 0.387 | 0.289 |
| 15.63 | 0.248 | 0.150 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-99% | 95% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 90-99% | 85-97% | 79-101% |
| EDTA plasma(n=5) | 86-97% | 93-102% | 87-98% | 83-97% |
| Heparin plasma(n=5) | 95-104% | 82-96% | 92-103% | 90-99% |