| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.157 | 2.076 |
| 2500.00 | 1.657 | 1.576 |
| 1250.00 | 1.174 | 1.093 |
| 625.00 | 0.873 | 0.792 |
| 312.50 | 0.528 | 0.447 |
| 156.25 | 0.423 | 0.342 |
| 78.13 | 0.248 | 0.167 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-107% | 100% |
| EDTA plasma(n=5) | 94-105% | 98% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 86-92% | 91-103% | 89-97% |
| EDTA plasma(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |
| Heparin plasma(n=5) | 85-97% | 87-96% | 82-95% | 92-105% |