| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.052 | 1.959 |
| 10.00 | 1.561 | 1.468 |
| 5.00 | 1.138 | 1.045 |
| 2.50 | 0.827 | 0.734 |
| 1.25 | 0.513 | 0.420 |
| 0.63 | 0.393 | 0.300 |
| 0.32 | 0.225 | 0.132 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 78-93% | 85% |
| Heparin plasma(n=5) | 78-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 86-98% | 92-103% | 87-93% |
| EDTA plasma(n=5) | 79-91% | 89-103% | 82-95% | 86-99% |
| Heparin plasma(n=5) | 86-97% | 90-98% | 80-93% | 92-105% |